Product Details:
| Minimum Order Quantity | 1 Piece |
| Model Name/Number | S- 935/ Microprocessor Flame Photometer |
| Brand | SYSTONIC |
| Automation Grade | Automatic |
| I Deal In | New Only |
| Capacity | 0.5 ml. |
| Weight | 7.5 KG (Approx.) |
| Sensitivity | Na: 0.5 ppm, K: 0.5 ppm |
| Product Type | Flame photo meter |
| Resolution | 0.1 |
| Power Supply | Smps Power Supply |
| Displayed Resolution | Graphical LCD 128x64 |
| Model Number | S-935 |
| Ignition System | Auto Ignition System |
| Gas Control | Adjustable with knobs |
| Flame System | LPG and Dry Oil Free Air |
| Usage/ Application | Industrial |
| Detection | Silicon Photodiode |
| Curve Accuracy | <2% error when 3 ppm Na/K and 5 ppm, Li are set as maximum Standards |
| Warranty | 12 Months |
| Calibration | upto 5 Point With curve fitting software |
| Average Time | in-build Software |
| Atomizer | Axial Flow Type |
| Filters | Na & K Supplied (Ca & Ba are optional at extra cost) |
| Color | White |
| Portable | No |
| Quantity Per Pack | 1 |
| Condition | New |
| Dimensions | 290 x 255 x 210 mm (L x B x H) (Approx.) |
| Grade | Automatic |
| Material | aluminium |
| Power | 230 VAC + 10% AC,50 Hz |
Microprocessor Flame Photometers S-935
Are used for the determination of sodium, potassium, calcium and lithium. They uses the latest microcontroller technology and advanced engineering techniques so as to give enhanced and reproducibility. They have soft touch membrane key for ease of operations.
Technical Specifications:-
General
Range Na: 0 – 100 ppm
Upto 250 meq/l, 1:100 dil
K: 0 – 100 ppm
Upto 250 meq/l, 1:100 dil
Ca: 0 – 100 ppm
Upto 250 meq/l, 1:100 dil
Li: 0 – 100 ppm
Upto 250 meq/l, 1:100 dil
Ba : 50-1000
Serum
Na: 0 – 200
1:100 dil
K: 0 – 10
1:100 dil
–
Li: 0 – 2
1:5 dil
Urine
Na: 0 – 250
1:100 dil
K: 0 – 100
1:100 dil
Ca: 0 – 10
1:2 dil
–
Sensitivity-:- General: Na: 0.5 ppm, K: 0.5 ppm, Li: 0.5 ppm, Ca: 15 ppm, Ba : 50 ppm
Filters:-Na & K supplied (Ca, Li & Ba are optional at extra cost)
Resolution:- 0.1 ppm/meq
Reproducibility:-< 1% coefficient of variation for 20 consecutive samples using 10 ppm Na set as maximum standard
Curve Fit Accuracy:- < 2% error when 3 ppm Na/K and 5 ppm Li are set as maximum standards
Display:-Graphical LCD with backlit
Average Time:- In-built in software
Flame System:-LPG and dry oil free air
Detector:- Silicon photodiode
Calibration:-with up to 5 Point Calibration with Curve Fitting Software (5 for each Element)
Multiple Calibration Curves can be stored in the memory
Gas Control:- Adjustable with knobs
Flame Failure:- Auto detection
Gas Cuttoff:- Automatic
Ignition System:-Auto ignition
Automiser:-Axial flow type
Interface Port:- USB
Power:- 230 V ± 10% AC, 50 Hz
Dimensions:-285 x 255 x 210 mm (L x B x H) (Approx.)
Weight:- 6.5 Kg (Approx.)
COMPRESSOR UNIT
Additional Information: